ECOLOG-PRO 1NTR
ECOLOG-PRO 1NTR

ECOLOG-PRO 1NTR

802250

Regular priceCHF211.50
/

Operational Conditions
Application


  • Measurement Module with One Built-In NTC Temperature Probe
  • Use ECOLOG-PRO 1NTR in a temperature range between -35 °C and -10 °C
  • Source two pieces of Lithium-Metal Batteries (AAA)
  • Two alkaline batteries included
  • Production Calibration Certificate
  • Lithium-metal batteries are not included
  • Yearly elproCLOUD Software Licenses to be added here...

 

Product Type: Measurement Module
Typical Application Area:
Pharmacy & Hospital, Warehouse & Cold Room,  Lab & Equipment, Cryo Storage & Shipment, Clinical Trials, Food, Museum, Server Room
Measurement Parameters: Temperature
Measurement  Temperature Range:
  • -35 °C..+50 °C (with optional LiFe batteries)
  • -10 °C..+50 °C (with Alkaline batteries)
Measurement Accuracy:
  • ±1.0 °C in the range between -35.0 °C and -10.1 °C
  • ±0.5 °C in the range between -10.0 °C and -0.1 °C
  • ±0.4 °C in the range between 0.0 °C and +25.0 °C
  • ±0.8 °C in the range between +25.1 °C and +50.0 °C

Measurement Interval:

1, 2, 3, 5, 10, 15, 20, 30, 60 minutes; user programmable via software (elproCLOUD & elproMONITOR)
Note: Logging interval will impact battery life
Usage:
Multiple Use
Required Communication Module:
ECOLOG-PRO RBR
Software:
elproCLOUD & elproMONITOR
Measurement Capacity:
14'800 measurement values
Certificates:

Production validation and 3-point calibration certificate;
Additional calibration based on manufacturer calibration standards or ISO 17025 available upon request


Technical Specifications ECOLOG-PRO 1NTR